NEMA 94 PET Phantom (NU 2-1994)™

Home NEMA 94 PET Phantom (NU 2-1994)™

NEMA 94 PET Phantom (NU 2-1994)™

Model PET/NEMA-94/P

Main Features
  • The NEMA 94 PET Phantom (NU 2-1994)™ is designed in accordance with the recommendations by the National Electrical Manufacturers Association (NEMA) to standardize the measurement of performance of PET*
Main Applications
  • PET acceptance testing with NEMA standard
  • Evaluation of count rate, uniformity, scatter fraction, attenuation compensation, and scatter compensation of ECT systems
  • Research
Specifications
  • All clear material: PMMA
  • Cylinder outside height with lid: 229 mm
  • Cylinder outside height without lid: 216 mm
  • Cylinder outside diameter: 203 mm
  • Cylinder inside: diameter: 197 mm
  • Wall thickness: 3 mm
  • Teflon® Insert diameter: 51 mm
  • Fillable Insert outside height: ~ 203 mm
  • Fillable Insert inside height: ~ 185 mm
  • Fillable Insert outside diameter: ~ 51 mm
  • Fillable Insert Inside diameter: ~ 45 mm
  • Line Source diameter: ~ 1 mm
  • Line Source height: ~ 184 mm
Volume
  • Empty: 6.5 liters
  • W/3 inserts: ~ 4.2 liters
  • The volume of each Fillable Insert: ~ 260 mL
Shipping
  • Carton: 13” x 13“ x 13”
  • Weight: 10lbs.

Get Inquiry

    Get Social